The Excelitas Noblelight SD3651-06TJ replaces the DO 650TJ and DO650TJ deuterium lamps. It is now sold as part number 80055323 (which replaces 80007103 and PE C055-0505) and is equivalent to PerkinElmer B016-0917.
Attention: The lamp is no longer available with a timer. It will only be provide a without a timer.
Applications
The Excelitas (Heraeus) Noblelight B016-0917 deuterium lamp is compatible with the following:
We recommend re-use of Adapter P/N B0166354. This may be required for some models of the PerkinElmer Lambda 5, 7 and 9. Do not discard during lamp changes.
We also sell the halogen replacement lamp B011-4620 for the same equipment, pre-aligned in bracket to be compatible with this lamp.
Spectral distribution: 160 - 400 nm, line free below 360 nm
Light aperture size: 1.0 mm
Overall length: 96.5 +1.5/-2.0 mm
Envelope diameter: 30 ± 0.8 mm
Envelope material: synthetic silica
Ignition voltage: 350 V
Operating voltage: 80 ± 10 Vdc
Operating current: 300 mAdc
Pre-heating: 12 Vdc / 0.87 Adc
Noise: ≤0.05 % p-p @ 215 nm
Drift: ± 0.2 %/h @ 215 nm
Lifetime: 1000 hrs
Connection: 3-pin connector
Pre-aligned, with timer, ELP coating
Note: Heraeus Noblelight was acquired by Excelitas in 2024. Branding on packages may reflect either company until transition is complete.
Also known as B0160917, LD-PER160, TruLife™ 1.5 Deuterium Lamp, L6022728, SD 3651-06 TJ, DO 650TJ, DO650TJ, C055-0505, SD 3651-06TJ, and 80055323.